Run Planning with the BSSH Run Planner Tool
Last updated
Was this helpful?
Last updated
Was this helpful?
To plan a successful sequencing run, a with details on run configuration (e.g., sequencer type, flowcell, and sample type) is required. Follow the instrument-specific steps below to create a sample sheet compatible with the Illumina Prep Kit and DRAGEN Protein Quantification.
Log in to and select your workgroup.
In the Run Planning tool, configure the settings described in the following table. Some settings are instrument specific. When you select the DRAGEN Protein Quantification application, the library prep kit and index adapter kit populate automatically, along with additional instrument-specific settings.
This table describes the possible configuration settings and values.
Instrument Platform
NovaSeq 6000/6000Dx
Secondary Analysis
[Cloud analysis] BaseSpace/Illumina Connected Analytics
[Local analysis] Local
Application
DRAGEN Protein Quantification (select the latest version)
Library Prep Kit
Illumina Protein Prep 9k (auto-populated)
Index Adapter Kit
Illumina DNA-RNA UD Indexes Set A B C D Tagmentation (auto-populated)
[NovaSeq 6000]
These settings are configured automatically and are not editable. - 2 indexes - Single Read - 15, 10, 10, 0
Lane Splitting is not supported for NovaSeq 6000 runs. Select "Repeat set of samples across all lanes."
Upload Illumina Protein Prep Automation System output file (*.csv) to BSSH Run Planner.
Select Import samples, select the CSV file type, and upload the Illumina Protein Prep Automation System output file. The interface highlights invalid values immediately after the file is rendered.
[Optional] To include a second plate in the run, repeat the import process and select Add to existing samples when prompted. The new samples are appended to the samples that were uploaded previously.
WARNING - Sample IDs and index sequences must be unique within a sequencing run. If combining libraries from multiple Illumina Protein Prep runs, avoid combining plates that contain the same sample IDs or index sequences.
[Optional] Multi-project analysis: Users may add Project
values either in the Illumina Protein Prep output file (prior to uploading) or add values in the BSSH Run Planner. See for more details.
WARNING - All samples from the same plate must have the same project (or no project) value.
[Optional] Enter an appropriate output file prefix. This value is used as a part of the prefix for the secondary analysis output file names. The first character must be alphanumeric. For the remaining characters, alphanumeric, hyphens, underscores, and spaces are permitted.
Proceed to the Run Review page and save the planned run.
[NovaSeq 6000, cloud analysis] Download the sample sheet and save it to a network location accessible to the sequencing instrument.
[NovaSeq 6000 Local analysis] Select Export to download the sample sheet. Save the file to a network location accessible to the sequencing instrument.
For additional information on run planning, refer to the documentation.
Log in to and select your workgroup.
If lane splitting will be utilized with this sequencing run, it's recommended to add values to Lane
column to Illumina Protein Prep Automation System output files locally. See for more details.
[Optional] Multi-project analysis: Users may add Project
values either in the Illumina Protein Prep output file (prior to uploading) or add values in the BaseSpace Run Planner. See for more details.
WARNING - All samples from the same plate must have the same project (or no project) value.
For additional information on run planning, refer to the documentation.