Known Limitations
Known Limitations/Issues with DRAGEN Protein Quantification v3.0 Software:
Only supports 96 samples per PlateBarcode for a single analysis.
SampleIDs must be unique prior to uploading SomaSeq Automation System output files to BaseSpace Sequence Hub run planner tool.
Calibrators provided with MultiMatrix runs are retained in the ADAT but are not used by autolaunch normalization. Explore DRAGEN Protein Quantification: Advanced Normalization Options in ICM to use matched sample-type calibrators.
Runs from different assay versions (v1 and v2) of SomaSeq Discovery cannot be combined in the same analysis.
If multiple kit types are present in the same flowcell, separate ADATs are generated for each kit type.
DRAGEN Protein Quantification v3.0 only supports the following Probe Pool lot combinations: CSF and MultiMatrix kits require Probe Pool AA; Plasma and Serum kits require Probe Pool AB.
DRAGEN Report specification page for MultiMatrix incorrectly displays the specification range for QC Check (PlatformSpecificQCCheckTailPercent) as 0.8 - 1.2; the correct range applied by the software is 0.4 - 1.6.
Known Limitations/Issues with Local Sample Sheet Generation Tool v3.0:
Sample sheet must be saved as "CSV Comma-Delimited" and not "CSV UTF-8".
Tool only supports starting from SomaSeq Automation System File v2, supporting SomaSeq Discovery v2 Assay.